Nebulizer, MicroFlow, perfluoroalkoxy (PFA), with Agilent integrated autosampler(I-AS) probe, flow rate 200 µL/min, standard with Agilent 7700s, 8800 ICP-MS in semiconductor configuration and 7500s ICP-MS, for analysis of high purity and hydrofluoric acid samples when using integrated autosampler, sample uptake tube terminates as a sampling probe
Agilent's high quality ICP-MS Nebulizers disperse the measured sample solution from the peristaltic pump as a fine mist into the spray chamber. This is carried out using a stream of argon gas. There are three types of nebulizers available: Babington nebulizers for introducing samples with high matrix and high TDS (total dissolved salts) levels, concentric nebulizers for use with clean samples (available in Pyrex or quartz versions) and CrossFlow nebulizers which are the standard nebulizers for the ICP-MS inert kit. A full range of connection accessories including tubing, end caps, O-rings and other fittings are available for all models.
- Unit of Measure:
- EA
- Spectroscopy Instrument Compatibility:
- 7700;7800;7850;7900;8800;8900
- Chemical Compatibility:
- Aqueous;Organic solvents;Strong acids;HF acid
- Autosampler Compatibility:
- I-AS
- Nebulizer Gas Connector:
- Threaded connector
- Nebulizer Material:
- PFA
- Nebulizer Name:
- MicroFlow
- Nebulizer Sample Particle Size Tolerance:
- 40 um
- Nebulizer Type:
- Concentric
- Notes:
- Standard for semicon configuration on 7700s, 8800, 8900, and 7500s For analysis of high purity and HF samples when using the integrated autosampler (I-AS), sample uptake tube terminates as a sampling probe
- Sample Connection:
- Direct
- Sample Uptake Rate:
- 200
- Self Aspirating:
- Yes, 2000 uL